Bilgilendirme: Kurulum ve veri kapsamındaki çalışmalar devam etmektedir. Göstereceğiniz anlayış için teşekkür ederiz.
 

Electro- Acoustic Device for Hip Dysplasia Assessment

No Thumbnail Available

Date

2016

Journal Title

Journal ISSN

Volume Title

Publisher

SYSCOM 18 S.R.L.

Open Access Color

OpenAIRE Downloads

OpenAIRE Views

Research Projects

Journal Issue

Abstract

A device for making diagnosis of dysplasia at the development fracture in newborns, assessment of osteoporosis and injuries of the skeletal system is presented. Its functioning is based on generation of acoustic resonance by sound transmitted through the bone under study. The device operates with a transmitter and an acoustic receiver coupled to the surface, just above the bone area under study. The measurements at the femoral bone in newborns indicate that the dominant frequency is around 160 Hz, which is consistent with other studies. Data comparisons with ultrasound technique suggest that this device could be an alternative for both dysplasia's studies of the hipbone and estimations of bone density.

Description

Gomez-Aguilar, J.F./0000-0001-9403-3767; Cordova-Fraga, Teodoro/0000-0002-6486-7530

Keywords

Acoustic Resonance, Densitometry, Hip Dysplasia, Nondestructive Technique, Osteoporosis

Turkish CoHE Thesis Center URL

Fields of Science

Citation

WoS Q

N/A

Scopus Q

N/A

Source

Revista de Chimie

Volume

67

Issue

6

Start Page

1140

End Page

1143
Google Scholar Logo
Google Scholar™

Sustainable Development Goals

4

QUALITY EDUCATION
QUALITY EDUCATION Logo

9

INDUSTRY, INNOVATION AND INFRASTRUCTURE
INDUSTRY, INNOVATION AND INFRASTRUCTURE Logo